• Accredited Company

    ISO 17025:2017 / ISO 9001:2015

10-50kW Power Instrument

Nominal Impedance 50Ω
Maximum Forward Power2 50kW
Forward Power Accuracy1 (Power Reading/10kW)% or 0.01% of the Max Forward Power, whichever is higher
Calibration Report
NIST Traceable, ISO17025:2017 Compliant
Maximum Reflected Power2 10kW
Reflected Power Accuracy1 Same as Forward Power
Nominal Frequency 13.56MHz
Frequency Range4 ±5% of Nominal Frequency
MAX Power Allowable 120% of Nominal Power
Directivity Min. 30dB
Max. Allowable Terminating VSWR Continuous use: VSWR is restricted to any load that requires more than 34.6Arms
Calibration Technique Power calibration factors stored in nonvolatile memory
Calibration Cycle, Nominal 12 months
Insertion Loss Max. 0.05dB
Harmonic Rejection Min. 20dB for narrow band devices
Pulsing Signal Capability Min. 500Hz
Gamma Calculated from Pr – Reflected power and Pf - Forward power:
Gamma
VSWR Calculated from Gamma:
VSWR
Operating Power Wall plug adapter (standard), USB or battery
RF Connectors3 EIA 3-1/8 Female - both ends
Battery Li-Ion
Battery Life 12 Hours minimum
Charge Time 6 hours maximum
Battery Charger Built in. Power supplied by wall plug adapter or USB
Display LCD 4 lines x 16 characters
Interfaces RS-232 and USB 2.0
Output Data Rate, Nominal 15 readings/second
1- Accuracy is measured into a 50Ω pure resistive load
2- Forward and Reflected Channel power levels can be custom scaled up to 3kW
3- End connectors: different end connectors are available on request
4- Optional feature
Humidity, Max. 95% (non-condensing)
Operating Temperature Range 20°C to 35°C (+68°F to +95°F)
Altitude 78.8kPa to 106kPa (6,562ft/2,000m to -1,640ft/-500m))
   
Storage Relative Humidity, Max 95% (non-condensing)
Altitude Standard atmospheric pressure variations, not exceeding 523mbar - 3,000m/10,000ft
Storage Temperature Range -20°C to + 70°C (-4°F to +158°F)
Dimensions (without connectors) 6”L x 6”H x 12”L (150 x 150 x 300mm)
Weight, Nominal 10lb. 0oz. (4.5kg)