• Accredited Company

    ISO 17025:2017 / ISO 9001:2015

600W-3kW Power Instrument

Nominal Impedance 50Ω
Nominal Forward Power2 3,000W
Forward Power Accuracy1 0.5% of Reading or 1.5W, whichever is higher, at Nominal Frequency;
2.0% of Reading or 3.0W, whichever is higher, at other frequencies, different than Nominal Frequency
Calibration Report
NIST Traceable, ISO17025:2017 Compliant
Reflected Power Accuracy1 1% of Reading or 3.0W, whichever is higher, at Nominal Frequency;
2% of Reading or 3.0W, whichever is higher, at other frequencies, different than Nominal Frequency
Nominal Frequency 13.56MHz
Frequency Range4 ±5% of Nominal Frequency
MAX Power Allowable 120% of Nominal Power
Directivity Min. 30dB
Max. Allowable Terminating VSWR Continuous use: VSWR is restricted to any load that requires more than 8.5Arms
Calibration Technique • Power calibration factors stored in nonvolatile memory, performed at minimum 3 Power Levels
• Temperature compensation
• Frequency compensation
Calibration Cycle, Nominal 6 months
Insertion Loss Max. 0.02dB
Harmonic Rejection Min. 20dB
Pulsing Signal Capability Min. 500Hz
Gamma Calculated from Pr – Reflected power and Pf - Forward power:
Gamma
VSWR Calculated from Gamma:
VSWR
Operating Power Wall plug adapter (standard), USB or battery
RF Connectors3 HN Female - both ends
Battery Li-Ion
Battery Life / Charge Time 12 Hours minimum / 6 hours maximum
Battery Charger Built in. Power supplied by wall plug adapter or USB
Display LCD 4 lines x 16 characters
Interfaces RS-232 and USB 2.0
Output Data Rate, Nominal 15 readings/second
1- Accuracy is measured into a 50Ω pure resistive load
2- Forward and Reflected Channel power levels can be custom scaled up to 3kW
3- End connectors: different end connectors are available on request
4- Optional feature
Humidity, Max. 95% (non-condensing)
Operating Temperature Range 20°C to 35°C (+68°F to +95°F)
Altitude 78.8kPa to 106kPa (6,562 ft/2,000m to -1,640ft/-500m)
   
Storage Relative Humidity, Max 95% (non-condensing)
Altitude Standard atmospheric pressure variations, not exceeding 523mbar - 3,000m/10,000ft
Storage Temperature Range -20°C to + 70°C (-4°F to +158°F)
Dimensions (without connectors) 2.5”L x 3.75”H x 4.75”L (63 x 95 x 121mm)
Weight, Nominal 2lb. 4oz. (1.02kg)